AGFA IMPAC CV DICOMStore with Media Purge Daemon (MPD) and IMPAX CV DICOMStore with Cardiovascular Purge Services (CPS), Medical Image Storage Device
FDA Recall Z-2252-2012 · Posted 2012-08-22
Advertisement
Recalling FirmAGFA Corp.
Recall StatusTerminated
Product CodeLLZ
Date Initiated2012-08-09
Date Terminated2015-12-08
Product Quantity535
Reason for Recall
Loss of patient data can occur under certain circumstances due to misconfiguration of DICOMstore configured with Agfa's Medical Image Storage Devices - Media Purge Daemon (MPD) and Cardiovascular Purge Service (CPS).
Firm Action
On 8/9/2012 AGFA Healthcare issued an "URGENT FIELD SAFETY NOTICE" letter to consignees. This letter describes the problem and mitigation. Acknowledgment, via FAX-Back or email, that the information was received and understood was also required from the consignees
Root Cause
Process design
Distribution Pattern
USA Nationwide, Virgin Islands and Canada