OxySure Model 615 disposable replacement cartridges, for portable oxygen generator.
FDA Recall Z-2253-2015 · Posted 2015-07-24
Advertisement
Recalling FirmOxysure Systems, Inc.
Recall StatusTerminated
Product CodeCAW
Date Initiated2015-06-12
Date Terminated2020-10-19
Product Quantity1005 units
Reason for Recall
Flow rate of the recalled replacement cartridges does not meet the draft FDA standard of a minimum 6 LPM for a minimum of 15 minutes.
Firm Action
The recalling firm will notify their affected customers via email beginning the week of 6/15/15. The recall notification will have a response for to be completed by the consignee. The recall notification calls for the customers to dispose of the affected product according to the MSDS.
Root Cause
Under Investigation by firm
Distribution Pattern
Nationwide throughout the US and Chile, Trinidad, China
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.