Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-3X6 G48278; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of biliary stones and foreign bodies.

FDA Recall Z-2253-2023 · Posted 2023-07-28

Recall StatusOpen, Classified
Product CodeFFL
Date Initiated2023-06-22
Product Quantity402 units

Reason for Recall

Non-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from handle (with or without mechanical lithotripsy).

Firm Action

Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice to their consignees on 07/14/2023 by email or priority mail. The notice explained the issue, the risk, and requested the consignee cease use and return the affected product.

Root Cause

Process control

Distribution Pattern

Worldwide distribution.