Terminated

Computed Tomography X-ray Systems Brilliance CT 64-channel with Essence technology. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

FDA Recall Z-2254-2015 · Posted 2015-07-24

Recall StatusTerminated
Product CodeJAK
Date Initiated2015-02-05
Date Terminated2018-06-26
Product Quantity4 Units

Reason for Recall

5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict

Firm Action

On 2/5/2015 the firm sent Customer Information Letters to their customers.

Root Cause

Software change control

Distribution Pattern

Worldwide Distribution-US (nationwide) including the states of AR, AZ, CA, CT, DE, FL, GA, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VT, WV & WY, and to the following countries: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cambodia, Canada, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Italy, Japan, Latvia, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Norway, Panama, Philippines, Russia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Viet Nam.