Terminated

Optima XR220amx Mobile general purpose radiographic imaging of the human head and body.

FDA Recall Z-2255-2012 · Posted 2012-09-13

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeIZL
Date Initiated2012-07-10
Date Terminated2013-01-10
Product Quantity254 total in the US

Reason for Recall

Software issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure without the technique factors being displayed.

Firm Action

The firm will issue Important Electronic Product Radiation Warning letters that identify the affected product along with a description of the defect and related hazards. Users are instructed to ensure that the acquisition mode is always selected prior to attempting an exposure to confirm the techniques selected. The firm will remedy the issue or bring the device into compliance free of charge. A GE Healthcare Service Representative will update the software on the system to address the issue. Field Modification Instructions (FMI) 10865(2) describes the rework plan the GEHC Engineers will make to the imaging system to return it to compliance with the applicable performance standard. It appears to adequately address the problem and is hereby approved. Users are to contact 800-437-1171 in the US, if they have any questions or concerns.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

Worldwide Distribution -- USA (nationwide) Distribution