Tibila Insert CK, Balanced Knee System Revision, REF563-1YYY Product Usage: UHMWPE component used in TKA and revision TKA procedures to serve as a tibial bearing surface for articulation with the femoral component where additional joint constraint is needed

FDA Recall Z-2257-2014 · Posted 2014-08-17

Recall StatusTerminated
Product CodeJWH
Date Initiated2014-07-11
Date Terminated2014-10-06
Product Quantity1

Reason for Recall

Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.

Firm Action

Ortho Development notified all affected consignees/distributors of the recall via email on July 11, 2014 which included a list of the affected product (part numbers, lot numbers and quantities) under their control. An RMA number and return shipping label were also included with the email. Emails were followed up with phone calls to verify receipt of the notification.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.