ARTIS icono systems with Quantification Application SW (QVA/QCA) on a standalone XWP computer, an interventional Fluoroscopic X-Ray System Model Number: 11327600

FDA Recall Z-2257-2020 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2020-04-23
Date Terminated2020-10-26
Product Quantity5 US

Reason for Recall

Using Quantification Application (QVA/QCA) on DSA images may lead to a failure in vessel detection or incorrect quantification of vessel detection and result in an incorrect diagnosis and inappropriate treatment of the patient

Firm Action

Siemens issued Urgent Medical Device Correction (AX035/20/S) dated 4/23/20. The letter states reason for recall health risk and action to take: We urgently recommend not to use Quantification Application (QVA/QCA) on DSA images until the software problem has been resolved. If Quantification Application has already been used on DSA images in the past, please verify the results and diagnostic evaluation if applicable. Siemens will correct the error with a software patch via Update Instruction AX034/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.

Root Cause

Software design

Distribution Pattern

AZ, IL, UT, WI