VisuALIF,System consisting of a 'U' shaped PEEK block in two footprint configurations, four heights and three lordosis angles., part number: 0301-1410. Intended to be implanted via an open anterior approach.

FDA Recall Z-2258-2013 · Posted 2013-09-19

Recall StatusTerminated
Product CodeMAX
Date Initiated2013-07-31
Date Terminated2014-03-18
Product Quantity32

Reason for Recall

The product has the potential to be laser marked as a Medium when it actually is a Small.

Firm Action

The firm decided to remove this lot and verify all of the markings. A stock sweep for mis-marked parts was performed at the time and it is not known why this lot is identified, however an additional stock sweep of all PEEK products from all product lines was performed beginning 07/31/2013 and no additional lots were identified.

Root Cause

Process control

Distribution Pattern

Distributed in FL.