GEHC OEC Injection Cable for MedRad Provis V Contrast Injector. In certain fluoroscope X-ray modes, the cable provides a signal to the auto-injector to trigger a contrast injection. Part numbers include 00-902260-01 and 00-883654-01.

FDA Recall Z-2262-2012 · Posted 2012-08-23

Recall StatusTerminated
Product CodeOXO
Date Initiated2008-10-10
Date Terminated2013-01-10
Product Quantity149 units

Reason for Recall

GE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.

Firm Action

An Urgent Safety Advisory Notice was sent to consignees on October 10, 2008. The letter identified the affected product and notified customers of the situation, along with instructing them to discontinue use until proper validation and instruction update for cable use was completed. GEHC OEC was to have contacted customers when the solution became available and resolved it at no cost to the customer.

Root Cause

Device Design

Distribution Pattern

Nationwide distribution.