QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000. The QUADROX oxygenators are intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.
FDA Recall Z-2262-2013 · Posted 2013-09-20
Advertisement
Recalling FirmMaquet Cardiovascular Us Sales, Llc
Recall StatusTerminated
Product CodeDTZ
Date Initiated2013-08-15
Date Terminated2017-07-11
Product Quantity1,182 units
Reason for Recall
The manufacturer has conducted a review of the results of product testing performed during the last twelve months and has identified specific lots of product which may not meet specifications for bond strength on the inlet and/or outlet connector.
Firm Action
Maquet sent a Urgent Product Recall letter/Response Form, dated 8/15/2013 via Fed Ex. On 8/26/2013 Maquet sent a letter to their customers informing them that all shipments of oxygenators and stainless steel hold clamps have been placed on a temporary hold due to an agency review in connection with FDA clearance. Customer service may be contacted at 1-888-627-8383, (option 2 followed by option 2) Monday though Friday between the hours of 8:00 am and 6:00 pm EST.
Maquet issued a Certificate of Medical Neccesity (CMN) on September 17, 2013 to notify Quadrox customers that they may continue to receive and use their current inventory of Quadrox oxygenators and clamps by acknowledging the conditions provided in the CMN.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution, including Nationwide (US) and foreign countries.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K112360QUADROX-I MICROPORPOROUS MEMBRANE OXGENATOR SERIES AND QUDROX-ID DIFFUSION MEMBRANE OXYGENATORS SERIESMaquet Cardiopulmonary, AG · 2011-09-15K103191QUADROX-I PEDIATRIC MICROPOROUS MEMBRANE OXYGENATOR WITH INTEGRATED ARTERIAL FILTER WITH SOFTLINE / BIOLINE COATINGMaquet Cardiopulmonary, AG · 2011-03-08K093522QUADROX-I SMALL ADULT MICROPOROUS MEMBRANE OXYGENATOR; W/ BIOLINE COATING AND W/ INTEGRATED ARTERIAL FILTER W/ BIOLINE CMaquet Cardiopulmonary, AG · 2009-12-07K090689QUADROX-I SMALL ADULT WITH AND WITHOUT INTERGRATED ARTERIAL FILTER WITH SOFTLINE COATING, MODELS HMO 50000, HMO 51000Maquet Cardiopulmonary, AG · 2009-10-29K090511QUADROX-I ADULT MICROPOROUS MEMBRANE OXYGENATOR WITH BIOLINE COATING, MODEL BEHMO70000,BEQ-HMO70000Maquet Cardiopulmonary, AG · 2009-06-12K082117QUADROX-I ADULT WITH AND WITHOUT INTEGRATED ARTERIAL FILTER WITH SOFTLINE COATING, MODEL# HMO 70000, HMO 71000Maquet Cardiopulmonary, AG · 2009-02-18