Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.

FDA Recall Z-2262-2026 · Posted 2026-05-29

Recall StatusOpen, Classified
Product CodeNEP
Date Initiated2026-04-22
Product Quantity105675

Reason for Recall

Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.

Firm Action

On April 22, 2026, the firm began notifying consignees via letters titled URGENT PRODUCT RECALL. Customers were instructed to locate impacted product. Customers may choose to inspect impacted inlets and continue to use those with the correct 3-micron filter, or they may choose to discard impacted inlet lots without inspection and product will be replaced.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, and West Virginia. The countries of Taiwan and Uruguay.