Nitinol Staple, Elastic Staple Small-Symmetrical, Sterile R, Model #: a) ES-08x08; b) ES-10x10; c) ES-10x13; d) ES-12x10; e) ES-15x12; f) ES-10x14; g) ES-10x15

FDA Recall Z-2265-2018 · Posted None

Recalling FirmTriMed Inc.
Recall StatusTerminated
Product CodeJDR
Date Initiated2018-04-13
Date Terminated2021-01-22
Product Quantity1,194

Reason for Recall

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Firm Action

On 04/13/18, customer recall communication forms were sent via mail informing customers to quarantine effected devices, call the manufacturer to arrange device return at 800-633-7221, and report adverse events to the FDA's MedWatch.

Root Cause

Process design

Distribution Pattern

U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden