Terminated

IntelePACS software versions 3-8-1 to 4-9-1, Product Usage: IntelePACS is a device that receives digital images and data from various sources (such as, CT scanners, MRI scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources.

FDA Recall Z-2267-2014 · Posted 2014-08-19

Recall StatusTerminated
Product CodeLLZ
Date Initiated2014-07-03
Date Terminated2016-01-13
Product Quantity230

Reason for Recall

In specific situation a partially truncated file may be written to the IntelePACS server. This partially truncated file still can be viewed by InteleViewer and InteleViewer will report to the user that the file in question is complete. This situation only occurs if a file is partially truncated.

Firm Action

Intelerad sent an Urgent Recall for Product Correction letter to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that a new software release will be installed on the system. Afterwards, an audit will be run on the the entire PACS to identify all image data that may have been truncated. Customers were requested to complete and return the attached Client Response Form.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - USA Nationwide and the countries of Scotland, Ireland, United Kingdom, Canada, Australia, and New Zealand.