Brand Name: B. BRAUN MEDICAL INC. Product Name: CESK RIVERSIDE HEALTH SYSTEM Model/Catalog Number: 552126 Software Version: N/A Product Description: CESK RIVERSIDE HEALTH SYSTEM Component: No
FDA Recall Z-2268-2026 · Posted 2026-05-29
Recalling FirmB Braun Medical Inc
Recall StatusOpen, Classified
Product CodeOGE
Date Initiated2026-01-29
Product Quantity690
Reason for Recall
B. Braun was notified by our supplier, Ferndale Laboratories, Inc., the manufacturer of MASTISOL Liquid Adhesive, that they have issued a voluntary recall due to potential cracking of the butyrate tubing (outside tube), which renders the product unusable.
Firm Action
On January 29, 2026 and May 6, 2026 URGENT MEDICAL DEVICE RECALL letters were sent to customers to notify them of a recall of a kit component. Actions to be taken include to review inventory to identify affected product, discontinue use and segregate. Using the acknowledgement form, respond to B Braun who will send further instructions to return the product. Should there be any questions, comments, or concerns regarding the product you have received, please feel free to contact (844-903-6417) or [email protected];
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of AL, FL, PA, TX, VA, WA.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.