Olympus JF-160F duodenoscopes (Model NumberJF-160F)
FDA Recall Z-2270-2018 · Posted None
Advertisement
Recalling FirmOlympus Corporation of the Americas
Recall StatusTerminated
Product CodeFET
Date Initiated2018-05-11
Date Terminated2020-05-05
Product Quantity68
Reason for Recall
Issuance of validated, new reprocessing procedures. The new cleaning procedure requires additional recess flushing and forceps elevator raising/lowering steps during precleaning, manual cleaning and manual disinfection.
Firm Action
Firm is extending recall to the user level
Letters will be sent on May, 11, 2018 via overnight mail to customers
Follow-up phone calls to consignees to offer training on new reprocessing instructions
No action can be taken with hospitals that are no longer in business
Root Cause
Under Investigation by firm
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.