Arrow Percutaneous Sheath Introducer Kit with Integral Hemostasis Valve/Side Port for use with 7.5 Fr. Catheters; Catalog number AJ-09801
FDA Recall Z-2274-2010 · Posted 2010-08-25
Advertisement
Recalling FirmArrow International Inc
Recall StatusTerminated
Product CodeKGZ
Date Initiated2010-06-07
Date Terminated2012-09-29
Product Quantity3000 kits
Reason for Recall
Sterility not assured. The stopcock handle position was inadvertently changed from an open to a closed position, which prevents sterilization of the closed portion of the device. Sterility of the affected product cannot be guaranteed.
Firm Action
June 7, 2010 Notification letter to customer instructed them to return the product to Arrow. Arrow will destroy the returned product.
Root Cause
Process change control
Distribution Pattern
All affected product was distributed to one foreign consignee: Arrow Japan Ltd.