Pressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter, Sharps Safety Features, and Maximal Barrier Precautions The ARROWg+ard Blue Plus" antimicrobial catheter is indicated in the short-term ( < 30 days) treatment of diseases or conditions requiring central venous access including, but not limited to the following: multiple infusions of fluids, medications, or chemotherapy; infusion of fluids that are hypertonic, hyperosmolar, or have divergent pH values; frequent blood sampling or blodd/blood component infusions; infusion of incompatible medications; central venous pressure monitoring; lack of usable peripheral IV sites; replacement of multiple peripheral sites for IV access

FDA Recall Z-2276-2017 · Posted None

Recall StatusTerminated
Product CodeFOZ
Date Initiated2017-03-20
Date Terminated2018-12-18
Product Quantityunknown-firm reports total devices distributed 43,394 each

Reason for Recall

Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.

Firm Action

Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to [email protected]. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide Distribution