Terminated

WS80A Diagnostic Ultrasound System Version 1.00.16, 3.00.17, 3.00.18, 3.00.19, 3.01.00, 3.01.01, 3.01.02, 3.01.03, 3.01.04, 3.01.05, 3.01.06, 3.01.07, 3.01.08, 3.01.09, 4.00.08, 4.00.09, 4.00.10, 4.00.11, 4.01.00, 4.01.01, 4.01.02, 4.01.03, 4.01.04, 4.01.05, 4.01.06, 4.01.07, 4.01.08, 4.01.09.

FDA Recall Z-2276-2019 · Posted None

Recall StatusTerminated
Product CodeIYN
Date Initiated2019-07-25
Date Terminated2020-09-11
Product Quantity5088 worldwide

Reason for Recall

There is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode.

Firm Action

On July 25, 2019, Neurological Corporation a subsidiary of Samsung Electronics issued Urgent Field Safety Notices to customers. Customers were advised to the following actions: 1) If the user wants to decrease the Dopper SV (Sample Volume) size set by himself/herself in the Doppler Only Mode, the user should press "set" button to update the PW Doppler mode. 2) A copy of the Field Safety Notice should be passed on to all who need to be aware within your organization. 3) Contact your local Samsung representative if you have questions. 4) Return the completed Customer Reply Form to Tom Leinart [email protected].

Root Cause

Software design

Distribution Pattern

US Nationwide