SunTech Medical Disposable, Neonate #4, 1-Tube, Male Bayonet, Blood Pressure Cuffs, Box of 20
FDA Recall Z-2277-2019 · Posted None
Advertisement
Recalling FirmSuntech Medical, Inc.
Recall StatusTerminated
Product CodeDXQ
Date Initiated2019-06-03
Date Terminated2020-12-16
Product Quantity80 total cuffs affected; 4 Boxes of 20 cuffs
Reason for Recall
Product is mislabeled. The label on the package is for a size #4 cuff, but the incorrect size #3 cuff was placed in the package.
Firm Action
On June 3, 2019, SunTech Medical sent an Urgent Medical Device Customer Notification to their one consignees advising them of the issue and requesting that they completed the enclosed Recall Acknowledgement Form. This letter further request this consignee to conduct a sub-recall by using the provided Customer Template Letter and Recall Acknowledgement Template.
Root Cause
Labeling Change Control
Distribution Pattern
Pakistan
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.