ABC Bend-A-Beam Handpiece REF 134009 Handcontrol Malleable Handpiece, 9" (22.8 cm) QTY: 1 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13205 ASSEMBLED IN MEXICO (Pouch label) ABC Bend-A-Beam Handpiece REF 134006 Handcontrol Malleable Handpiece, 9" (22.8 cm) QTY: 10 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13205 ASSEMBLED IN MEXICO (Case label) Usage: Electrosurgical handpiece for use as monopolar electrosurgical accessory in conjunction with CONMED electrosurgical units equipped with argon beam coagulation.

FDA Recall Z-2278-2010 · Posted 2010-08-25

Recalling FirmConMed Electrosurgery
Recall StatusTerminated
Product CodeGEI
Date Initiated2008-12-18
Date Terminated2010-10-22
Product QuantityDom: 233 cs (2330 units); Int'l: 275 cs (2750 units)

Reason for Recall

The internal electrode can protrude in front of ceramic insulator at the tip of the ABC Handpiece used with Conmed Electrosurgical units.

Firm Action

An Urgent Device Recall notification was issued 12/18/2008, which explained the problem/risk associated with the device, the discontinuance of use and the return of the device. If you have any questions, please contact 800-552-0138, ext. 5225.

Root Cause

Process design

Distribution Pattern

Nationwide and International