PHADIA Varelisa Cardiolipin IgM Cardiolipin autoantibody immunological test system (IgM class antibodies)
FDA Recall Z-2285-2012 · Posted 2012-08-28
Advertisement
Recalling FirmPhadia US Inc
Recall StatusTerminated
Product CodeMID
Date Initiated2012-08-02
Date Terminated2012-12-12
Product Quantity837 kits (96 assays per kit), total of 80,352 individual assays
Reason for Recall
Customers were not following the labeled assay procedure. This was due to a letter issued to them in 2004 from Phadia GmbH, which indicated that customers could omit the pre-wash step. The omission of the pre-wash step, when combined with a change in material source in kit lot #0067, caused elevated results.
Firm Action
On 8/2/2012 an URGENT PRODUCT CORRECTION INFORMATION FOR PHADIA VARELISA CARDIOLIPIN IGM, Article Number 15696 letter was sent to all consignees the received the Direction for Use modification letter in 2004.
Root Cause
Labeling Change Control
Distribution Pattern
USA Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.