Sparrow Ascent Patient Controller, REF's: 11-110-K, 11-310-K, 11-320-K, 11-321-K, 11-310-K-28, 11-320-K-28, 11-321-K-28
FDA Recall Z-2287-2025 · Posted 2025-08-11
Advertisement
Recalling FirmSpark Biomedical Inc
Recall StatusOpen, Classified
Product CodePZR
Date Initiated2025-07-02
Product Quantity78 units
Reason for Recall
Due to manufacturing issue, neurostimulator device may experience "cable disconnect" error message and interruption to stimulation output.
Firm Action
On July 2, 2025, Spark Biomedical initiated a Urgent Medical Device Recall via phone. During the phone conversation, Spark Biomedical will ask consignees to take the following actions:
1. Spark Biomedical will replace all affected00 units at no cost. Even if the units appear to be working.
2. Current devices may continue to be used until they are replaced.
3. Please perform a check of the cable to patient controller connection before use. If this connection is loose or moves, please discontinue use of that patient controller device and quarantine that device until the replacement arrives.
4. Please notify all appropriate personnel at your facility that need to know about this recall.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.