ReliaMed Deluxe Portable Adjusting Table. Product is a medical table used on patients for massage and chiropractic treatment.

FDA Recall Z-2288-2010 · Posted 2010-08-26

Recalling FirmMDC Acquisition Co.
Recall StatusTerminated
Product CodeFPS
Date Initiated2010-04-23
Date Terminated2012-02-16
Product Quantity265 Tables

Reason for Recall

Two complaints were received of the vinyl-covered, padded, plywood table surface breaking near the foot end.

Firm Action

"Urgent: Medical Device Recall" letters were sent to consignees from Milliken Medical and Meyer Distributing Company in April 2010 describing the issue. They instructed their customers to immediately stop use of the table. customers were instructed to return the tables to the firm with an enclosed Recall Form for future replacement, or for replacement with a comparable table from another menufacturer at no cost, or for refund of the original cost of the device. Consignees may contact Meyer Distributing Company at 800-472-4221 and Milliken Medical at 800-532-1356.

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution: USA, including the states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA and WI, and the country of Canada.