Terminated

Elekta Unity, UDI 05060191071321, Image-guided radiation therapy system - Product Usage: is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

FDA Recall Z-2289-2020 · Posted 2020-06-05

Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeIYE
Date Initiated2020-04-30
Date Terminated2021-08-04
Product Quantity32 units

Reason for Recall

potential risk that Legionella may be present in the Unity machine room, specifically the heat exchanger condensation collection tray.

Firm Action

The firm notified the service engineers of the issue and the risk to health (exposure to legionella) by email on 04/30/2020. The firm followed with a letter disseminated on 05/12/2020 to customers and service engineers which discussed the hazard and Elekta's updated preventative maintenance.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of PA, WI, NY, TN, NJ, IA, TX, and the countries of Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Thailand, Turkey, UK.