syngo Dynamics version 9.0, Siemens Medical Solutions, USA, Inc., Ann Arbor, MI. The syngo Dynamics is a picture archiving and communications system intended for acceptance, transfer, display, storage, archiving and manipulation of digital medical images, including quantification and report generation. It is not intended to be used for the reading of mammography images.
FDA Recall Z-2293-2011 · Posted 2011-05-25
Advertisement
Recalling FirmSiemens Medical Solutions, USA, Inc
Recall StatusTerminated
Product CodeLLZ
Date Initiated2011-03-24
Date Terminated2012-07-10
Product Quantity29 devices
Reason for Recall
If one or more graphs are in a table on the report, and if any of those graphs are deleted, it may lead to reports failing to save and/or marking the study as read might leave the study in an unread state or might mark the study as read but leave the report with a "Preliminary" watermark displayed.
Firm Action
The firm, SIEMENS, sent a "Customer Safety Advisory Notice" letter dated March 24, 2011 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to place graphs in sections and not tables in the report; avoid deleting graphs from the reports in run-time and include this safety advisory notice in their operator's manual where it should remain until the service patch is applied. SIEMENS will be issuing a service patch to fix the defect. The letter states the patch will be available in the next few weeks.
If you have any questions, call +1 (734) 205-2400.
Root Cause
Software design
Distribution Pattern
Nationwide distribution: USA including states of: CA, CT, FL, ID, MD, MI, MN, NJ, NY and PA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.