Terminated

Endotine Transbleph 3.5 (CFD-080-0167)

FDA Recall Z-2294-2016 · Posted None

Recall StatusTerminated
Product CodeHWC
Date Initiated2016-05-06
Date Terminated2016-08-12
Product Quantity29

Reason for Recall

MicroAire has initiated a limited recall (29 Units)on the Endotine Transbleph 3.5 (CFD-080-0167), Lot # 285373, due to minor mislabeling ( the front label is correct, but the side label lists an incorrect part number, i.e. CFD-080-0177).

Firm Action

The firm is sending recall letters to the customers who have received the Endotine Transbleph 3.5 (CFD-080-0167), lot #285373. Letters sent by UPS to expedite the delivery.

Root Cause

Labeling Change Control

Distribution Pattern

The product was distributed in the United States and Sweden

Related FDA 510(k) Records