Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

HardyCHROM MRSA,REF; G307, LOT: 447694, A selective and differential chromogenic medium for the qualitative direct detection of nasal colonization by methicillin-resistant Staphylococcus aureus (MRSA). Not intended to diagnose infection of guide therapy. Further testing required for susceptibility or typing. Container Type: 15 x 100mm monoplate Packaged: 10 plates/sleeve STORAGE: 2-8 degree C on receipt, LIGHT SENSITIVE IVD UDI: (01)08165760201400 (10)447694 (17)191206

FDA Recall Z-2298-2020 · Posted 2020-06-09

Recalling FirmHardy Diagnostics
Recall StatusTerminated
Product CodeJSO
Date Initiated2019-11-14
Date Terminated2021-04-28
Product Quantity3,580 ea (358 kits, pack of 10 in each)

Reason for Recall

False Positive results due to the MSSA (methicillin-susceptible) strain Staphylococcus aureus (ATCC 29213) breakthrough when this strain should have been inhibited.

Firm Action

On 11/14/2019, the firm sent a notification of recall letter to customers, informing them of the performance issue of the device lot of the breakthrough on firm's differential chromogenic medium. The customers were advised to discard the remaining stock, and notify the Technical Services Department. If the customer would like replacement or credit, contact Technical Services Department or via e-mail at [email protected], or 800-266-2222 option 2 or ext. 5658.

Root Cause

Nonconforming Material/Component

Distribution Pattern

U.S. Distribution in the states of WI, CA, WA, IA, MT, UT, AZ, WY, KS, MO, OR, TX, VI, including PR and Guam. OUS: British Virgin Islands, Kenya,

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑