BD Pyxis MedStation ES REF: 323 Medication cabinet
FDA Recall Z-2298-2026 · Posted 2026-06-03
Advertisement
Recalling FirmCareFusion 303, Inc.
Recall StatusOpen, Classified
Product CodeBRY
Date Initiated2026-04-30
Product Quantity15,933
Reason for Recall
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
Firm Action
On April 30, 2026, CareFusion (subsidiary of Becton, Dickinson and Company) issued a Urgent Medical Device Recall Notification, via mail or Email. CareFusion ask consignees to take the following actions:
1. Dispensing cabinets can continue to be used as intended.
2. If you suspect that the Bio-ID Scanner is hot, please discontinue the use of Scanner and file a complaint.
3. Review and update facility policies and procedures related to situations where the Bio-ID Scanner is unavailable to mitigate delays in care.
4. Complete the attached Customer Response Form and return to the BD contact noted on the form.
Root Cause
Component design/selection
Distribution Pattern
US: Nationwide OUS: AE, AU, BE, BH, BR, CA, CL, ES, GB, IT, KW, MA, MX, NZ, PT, QA, SA, SG, TH, TW
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.