OmniDiagnost-Eleva X-ray system Angiographic and Diagnostic, Product Nos. 708026, 708027, 708028
FDA Recall Z-2300-2018 · Posted 2018-06-27
Advertisement
Recalling FirmPhilips Medical Systems Nederlands
Recall StatusTerminated
Product CodeJAA
Date Initiated2018-05-01
Date Terminated2020-06-08
Product Quantity508
Reason for Recall
The fixation of the upper and lower tilt actuator, of the Omnidiagnost system might break off and the table will start to rotate from 0 to + 90 /- 90 degree with high speed. This rotating movement can not be stopped by the user. This could potentially lead to the patient falling off the table. There is a risk of injury for the patient, user, or bystander.
Firm Action
On May 17, 2018, the firm issued Urgent Medical Device Correction letters to affected consignees. Customers were advised of the product issue.
Customers were instructed to do the following:
Stop using the system and call your Philips representative if any of the following situations occurs:
. the table height or tilt movement is not working.
. you notice a cracking sound or snapping sound different than the normal
sound during height or tilt movement.
. you experience a blockade in the table height or tilt movement during
system movements.
. the system is having an unexpected collision or the system had an
unexpected collision in the past 2 months.
. you have used a CPR stand that obstructed the system, or
. you notice any (other) unusual system behavior beyond the normal use
of the system.
Avoid collisions with the table and base stand of the system by ensuring that no
obstacles are placed around the system.
Customer shall ensure that all staff with access to the affected systems are
informed of the content of this Field Safety Notice.
A copy of this Field Safety Notice shall be placed together with the
documentation of the system until the system has been corrected by Philips.
The firm plans to correct all potentially affected products in the field by means of a field change order free of charge. Customers will be contacted by a local Philips representative to schedule the corrective action, which will start effective July 2018.
Customer with further questions may contact Philips Technical Support Line: 1-800-722-9377.
Root Cause
Device Design
Distribution Pattern
US nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement