Terminated

Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Intended to create and display two-dimensional and three-dimensional images of anatomy from a series of digitally acquired images.

FDA Recall Z-2301-2016 · Posted None

Recall StatusTerminated
Product CodeLLZ
Date Initiated2016-01-30
Date Terminated2019-04-08
Product Quantity109 sites have the affected versions

Reason for Recall

Studies coming over via telmed were missing patient's DOB, procedure, and referring physician.

Firm Action

The recalling firm issued Urgent: Medical Device Recall letters dated 1/27/2016 and 5/26/2016. The letter informed the customer of the affected product, the nature of the issue, the potential harm involved, and actions to be taken. The firm released a fix for the issue; and, customers were asked to complete and return the enclosed form by June 17, 2016. Questions can be directed to [email protected].

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- USA, New Zealand, and United Kingdom.