Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Poly stat Flu A & B Test Manufactured for Polymedco 510 Furnace Dock Road Cortlandt Manor, NY 10567 Tel 800-431-2123 914-739-5400 Fax 914-739-5890. The BioSign Flu A + B test is an in vitro rapid qualitative test that detects influenza type A and type B nucleoprotein antigens directly from nasal swab, nasopharyngeal swab and nasal pharyngeal aspirate/wash specimens obtained from patients with signs and symptoms of respiratory infection. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. A negative test result is presumptive and it recommended that these results be confirmed by viral culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use.

FDA Recall Z-2307-2012 · Posted 2012-08-31

Recall StatusTerminated
Product CodeGNX
Date Initiated2012-03-09
Date Terminated2014-10-14
Product Quantity103 kits (22 tests per kit)

Reason for Recall

There have been reports of dual positive (Flu A + Flu B positive) results being encountered.

Firm Action

OraSure Technologies, Inc. sent an "IMPORTANT NOTICE FOR ORASURE QUICKFLU RAPID FLU A + B TEST" letter dated March 13, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact the firm at 610-882-1820 for questions concerning this recall.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution-including the states of NJ, NY, and PA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑