Open, Classified

TITANIUM/FLUOROPLASTIC PISTON 0.6MM DIA X 3.75MM L Model/Catalog Number: 409-375 Product Description: Stapes Prosthesis

FDA Recall Z-2308-2026 · Posted 2026-06-04

Recalling FirmGrace Medical, Inc.
Recall StatusOpen, Classified
Product CodeETB
Date Initiated2026-04-03
Product Quantity7

Reason for Recall

Due to incorrect functional length on device labeling.

Firm Action

On 04/03/2026, the firm emailed a "RECALL NOTICE" to it single customer informing them that the firm determined the pistons were manufactured with an incorrect functional length. Devices are labeled as 3.75 mm but have an actual functional length of 4.00 mm. The customer was instructed to: " Immediately discontinue use and quarantine affected product " Return affected product using the provided RMA " If any devices have been used, provide the date(s) of surgery on the response form " Complete and return the enclosed Recall Response Form For questions - contact Grace Medical at 901-386-0990 or [email protected].

Root Cause

Process control

Distribution Pattern

U.S Nationwide distribution in the state of MD.