Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Zilver 518 Vascular Self Expanding Stent The Zilver Vascular Stent is intended for use as an adjunct to percutaneous transluminal angioplasty (PTA) in the treatment of symptomatic disease of iliac arteries up to 100 mm in length with a reference vessel diameter of 5 to 9 mm. Patients should be suitable candidates for PTA and / or stent treatment.

FDA Recall Z-2309-2012 · Posted 2012-09-01

Recalling FirmCook Ireland Ltd.
Recall StatusTerminated
Product CodeNIO
Date Initiated2012-01-05
Date Terminated2013-02-04
Product Quantity15 units of ZIV5-18-125-8-80 LOT# CF694429

Reason for Recall

Cook Ireland received a product complaint from Japan on December 22nd, 2011. The complainant reported that a 5 French Zilver Vascular Stent (part number ZIV5-18-125-8-80 from Lot# CF694429) was inserted into a sheath over a wire guide where it became lodged in the sheath hub. After further examination, the user determined that the product was a 6 French Zilver Vascular Stent (part number ZIV6-35

Firm Action

The firm, Cook Medical, sent a letter dated February 13, 2012 to foreign consignee. The letter described the product, problem and actions taken. The firm stated that re-training of the relevant personnel has been carried out; they will continue to monitor to identify potential emerging trends and further corrective and preventive action will be considered in an effort to avoid the recurrence of similar events. If you have any question contact in US: Customer Relations Manager at 800.346.2686.ext 2222 or email: [email protected]; OUS: Customer Quality Supervisor at +353 61 239343.

Root Cause

Packaging process control

Distribution Pattern

Worldwide distribution: USA (nationwide) and country of: Japan.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑