BD Vacutainer LH (Lithium Heparin) 102 I.U. Plus Blood Collection Tubes, Catalog Number 368886

FDA Recall Z-2310-2019 · Posted None

Recall StatusTerminated
Product CodeJKA
Date Initiated2019-06-12
Date Terminated2020-06-29
Product Quantity1,436,228,004 total

Reason for Recall

False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

Firm Action

Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution, including US Nationwide