Hudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management

FDA Recall Z-2311-2019 · Posted 2019-09-09

Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeBTR
Date Initiated2019-07-30
Date Terminated2022-02-15
Product Quantity7963 units

Reason for Recall

Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.

Firm Action

Teleflex sent an Urgent Medical Device Recall Notification letter dated July 30, 2019 to affected customers. The letter explains the problem and requests the consignee discontinue use and quarantine of any of the recalled products then to notify Teleflex. The firm is requesting the return of the affected products. For questions, contact your local sales representative or Customer Service at 1-866-396-2111.

Root Cause

Process control

Distribution Pattern

US Nationwide Distribution