GE SIGNA Architect AIR (China Only), Nuclear Magnetic Resonance Imaging System
FDA Recall Z-2312-2025 · Posted 2025-08-13
Advertisement
Recalling FirmGE Medical Systems, LLC
Recall StatusOpen, Classified
Product CodeLNH
Date Initiated2025-06-20
Product Quantity151 units
Reason for Recall
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
Firm Action
GE HealthCare issued an URGENT CORRECTION NOTICE to it consignees on 06/20/2025 via traceable means. The notice explained the safety issue and requested the following actions be taken:
"You can continue to use your MR system by following the instructions below:
In the undocked position, lock the table by pressing the pedal in the front of the table. Check to see that all casters are locked and that the table is secure. Contact your GE HealthCare representative for assistance if any of the casters do not lock.
Before each patient transfer verify that the caster locks are engaged. If you notice more than one caster is damaged or fails to lock properly, use particular caution as the table may move while in use.
No additional steps are required to continue to use your table in the docked position.
Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions."
"GE HealthCare will inspect and if necessary, correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction."
" if you have any questions for concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative."
Root Cause
Equipment maintenance
Distribution Pattern
Worldwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.