Penumbra Neuron Delivery Catheter 053, 115 cm/12 cm-- MP, manufactured by Penumbra Inc., Alameda, CA Percutaneous Catheter, indicated for the introduction of interventional devices into the peripheral, coronary and neuro vasculature.

FDA Recall Z-2313-2009 · Posted 2009-09-24

Recalling FirmPenumbra Inc.
Recall StatusTerminated
Product CodeDQY
Date Initiated2009-07-01
Date Terminated2010-11-30
Product Quantity37 units

Reason for Recall

Misbranding--Some products may be identified with the wrong ID band content 'Neuron 6F 070' instead of 'Neuron 6F delivery'

Firm Action

Firm initiated its recall on June 30, 2009. Consignees notified on July 6 and 7, with follow up contact on July 7, 2009. Contact Penumbra Inc. at 1.510.618.3223, if there are any questions.

Root Cause

Mixed-up of materials/components

Distribution Pattern

CA, FL, NC, MI, MO, VA, AZ, CT, IA, NY, and CO, Foreign: Germany and Australia.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.