Terminated

LapSac Introducer, Supplied sterilized by ethylene oxide gas in peel-open packages. Used to aid in the introduction of a LapSac during endoscopic surgical procedures.

FDA Recall Z-2314-2017 · Posted None

Recalling FirmCook Inc.
Recall StatusTerminated
Product CodeGCJ
Date Initiated2017-05-01
Date Terminated2019-03-26
Product Quantity327 lots, 288 units

Reason for Recall

COOK Medical is initiating a voluntary recall of multiple products because the reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.

Firm Action

The firm, Cook Medical, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 26, 2017 to its customers. Letters were mailed to customers via courier on April 28, 2017. The letter described the product, problem, and actions to be taken. The customers were instructed to: examine your inventory immediately to identify and quarantine affected products; complete and return the required Acknowledgment and Receipt form (even if you do not have affected product) via fax (812.339.7316) or email ([email protected]); return and discard any affected products, and Immediately report any adverse events to Cook Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email at [email protected]. Should you have any medical questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution - US (nationwide) and countries of: Australia, Austria, Belgium, Bahrain, Canada, China, Costa Rica, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Lithuania, Latvia, Malaysia, Netherlands, Romania, Russia, Saudi Arabia, Singapore, Spain, Switzerland, United Arab Emirates, and United Kingdom.