ZOLL R Series Defibrillator/Pacemaker/Monitor Used by qualified medical personnel for the purpose of converting ventricular fibrillation to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.

FDA Recall Z-2315-2009 · Posted 2009-09-29

Recall StatusTerminated
Product CodeMKJ
Date Initiated2009-08-31
Date Terminated2012-05-30
Product Quantity247 units

Reason for Recall

Failure of Device to display ECG could delay delivery of defibrillation or pacing therapy.

Firm Action

Zoll notified domestic consignees on August 31, 2009 and international consignees on September 1, 2009 by an Urgent Medical Device Correction letter. Customers were instructed to continue to perform daily tests per the recommendations provided in the R Series Operator's Guide. In the event of a failure, customers are to remove the device from service and contact biomedical or clinical engineering department per their facility's protocol. Units that operate as expected can remain in service and back-up monitor should be available in the event of a failure. Zoll's technical support department should be contacted at 1-800-348-9011 or +1-978-421-9460 to schedule the return of affected devices with the defect for replacement of the suspect component.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution -- USA, UK, Austria, Australia,, Canada, Switzerland, Germany, France, French Guiana, Jordan, Macedonia, Netherlands, Panama, Poland, Romania, and Saudi Arabia.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.