E.faecalis/OE PNA FISH Intended Use: E. faecalis/OE PNA FISH" (Cat. No. KT003) is a multicolor, qualitative nucleic acid hybridization assay intended for identification of Enterococcus faecalis and selected other enterococci (OE) from blood cultures. E. faecalis/OE PNA FISH provides rapid identification of E. faecalis and selected other enterococci (OE) on smears made from positive blood cultures and is compatible with all major blood culture media

FDA Recall Z-2315-2010 · Posted 2010-08-31

Recalling FirmAdvanDx, Inc.
Recall StatusTerminated
Product CodeOAH
Date Initiated2010-06-08
Date Terminated2011-06-06
Product Quantity541 units

Reason for Recall

Some strains of streptococus anginosus cross-reach with E. faecalis/OE PNA FISH to produce a false positive green signal.

Firm Action

The firm, AdvanDx, sent a letter dated June 8, 2010, and email notification on June 11, 2010, to all customers. The letter described the product, problem and action to be taken by customers. AdvanDx notified the customers that there was a change to the package insert under "limitations" to include the statement "some strains of Streptococcus anginosus produce a false green positive fluorescence due to sequence similarities and enclosed a package insert for all E. facealis/OE PNA FISH (KT003) kits. The customers were instructed to replace the Package Insert (revision B) included with any kits in their inventory with the enclosed Package Insert (revision C). Note: All shipments of E. faecalis/OE PNA FISH, as of June 14, 2010, will contain this new package insert. If you have any questions or requests for further information, please contact AdvanDx Technical Support at US (Toll Free): 1-800-376-0009 or email [email protected].

Root Cause

Device Design

Distribution Pattern

Worldwide distribution: USA and countries including Danmark, Finland and Sverige (European).