Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017
FDA Recall Z-2318-2017 · Posted 2017-05-15
Advertisement
Recalling FirmZimmer Biomet Spine Inc.
Recall StatusTerminated
Product CodeLYQ
Date Initiated2017-05-10
Date Terminated2018-05-25
Product Quantity497 units
Reason for Recall
Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.
Firm Action
Consignees will be notified on 5/15/17 via certified mail to return products.
Root Cause
Device Design
Distribution Pattern
Nationwide, military/VA/govt and foreign consignee (No Canadian consignees)