Response BandLoc 5.5/6.0 Tensioner Instrument The Tensioner Instrument (01-1102-1000) is utilized to allow the surgeon to apply a recommended tension to the band and hold it in place during spine manipulation and set screw tightening.

FDA Recall Z-2321-2017 · Posted None

Recalling FirmOrthoPediatrics Corp
Recall StatusTerminated
Product CodeOWI
Date Initiated2017-04-24
Date Terminated2017-09-18
Product Quantity82

Reason for Recall

Over-tensioning of the band resulting in damage to the band causing it to rupture

Firm Action

The firm, OrthoPediatrics Corp., contacted each distributor/consignee via phone on 4/24/2017, regarding the recall. The phone call informed Consignees of issue and actions to be taken. The consignees were instructed to return of devices in their control were issued; cease further distribution of the product and quarantine it. The firm state that product will be distributed after Rework to new revision. If you have any questions, contact V.P. Regulatory Affairs, at (574) 268-6379 or email: [email protected].

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution-US distribution to: MO, NY, KY, FL, NC and OH; and countries of: AUSTRALIA, FRANCE and SPAIN.