Volara System. Single Patient Use (SPU) Circuit 5kit for Volara and Single Patient Use (SPU) Circuit kit for Volara. Intermittent positive pressure breathing device.
FDA Recall Z-2321-2024 · Posted 2024-07-17
Advertisement
Recalling FirmBaxter Healthcare Corporation
Recall StatusOpen, Classified
Product CodeNHJ
Date Initiated2024-05-30
Product Quantity58,220 units
Reason for Recall
Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.
Firm Action
Baxter Healthcare notified customers on about 05/30/2024 via letter sent first-class mail. The letter instructed facilities to immediately locate, isolate, and cease all use of the affected lot numbers, return affected units on hand, and complete and return the acknowledgement form. If the product was further distributed, they were instructed to notify customers if product was further distributed.
The letter instructed patients to immediately locate, isolate, and cease all use of the affected lot numbers, return affected units on hand, however if they do not have other options in stock, in order to not delay therapy, you may continue to use with caution. Visually inspect the nebulizer port prior to therapy and assess for ventilator gas flow leaks for the duration of therapy. They were also instructed to complete and return the acknowledgement form.
Root Cause
Device Design
Distribution Pattern
Nationwide distribution. International distribution to Canada, France, and Germany.