Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;
FDA Recall Z-2321-2026 · Posted 2026-06-05
Advertisement
Recalling FirmFresenius Medical Care Holdings, Inc.
Recall StatusOpen, Classified
Product CodeKPO
Date Initiated2026-05-12
Product Quantity1,756,108 units
Reason for Recall
Given the increased number of leaks reported by
users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard
associated with fluid on the floor.
Firm Action
On 5/12/2026 an IMPORTANT MEDICAL DEVICE INFORMATION letters were sent to customers to inform them that manipulation of the luer-lock location during treatment preparation may result in fluid spillage onto the user and floor. This occurs prior to treatment and renders the product unusable. Direct contact with the fluid does not present any risk to the user as frangible manipulation only occurs after the peal seam has been broken and the two fluid compartments have been mixed. The potential risk associated with the affected Bicarby Dialysate product is a slip and fall hazard due to the presence of fluid on the floor.
Root Cause
Process design
Distribution Pattern
US Nationwide distribution in the states of NY, TN and TX.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K243786Bicarby Dialysate RFP-402 (RFP-402-G); Bicarby Dialysate RFP-400 (RFP-400-G); Bicarby Dialysate RFP-407 (RFP-407-G); Bicarby Dialysate RFP-401 (RFP-401-G); Bicarby Dialysate RFP-404 (RFP-404-G); Bicarby Dialysate RFP-456 (RFP-456-G); Ci-Ca Dialysate 2K (RFP-457-G); Ci-Ca Dialysate 4K (RFP-458-G)Fresenius Medical Care Renal Therapies Group, LLC · 2025-04-04K233950pureFLOW 402 (F00012067); pureFLOW 406 (F00012068); pureFLOW 401 (F00012069); pureFLOW 400 (F00012070)Fresenius Medical Care Renal Therapies Group, LLC · 2024-05-10