Philips DuraDiagnost X-ray systems: DuraDiagnost (R1.0, Product Number 712210) DuraDiagnost (R2.0, Product Number 712211) DuraDiagnost (R3.0, Product Number 712211)

FDA Recall Z-2322-2017 · Posted None

Recall StatusTerminated
Product CodeKPR
Date Initiated2017-05-31
Date Terminated2018-08-24
Product Quantity98 sytems

Reason for Recall

Tube arm assembly could fall down due to cracked welding joints

Firm Action

Philips Healthcare sent an URGENT - Field Safety Notice Medical Device Correction letter dated May 23, 2017, advising users of the problem and asking facilities to stop using the device until a Philips Engineer can go on-site to perform the FC071200173; contact your local Philips representative if you have any concern or support required on this issue. If you need any further information or support concerning this issue, contact your local Philips representative. For further questions, please call (978) 659-3000.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US Distribution to the states of : Ca and Oh., and to the countries of: Australia, Libya, China, Mayotte, Mexico, France, United Kingdom, Germany, Ecuador, Australia, Spain, Switzerland, Macedonia, Uzbekistan, Kuwait, Malaysia, Netherlands, Argentina, Chile, Colombia, Egypt, United Arab Emirates, Saudi Arabia, Hungary, Peru, Poland, Indonesia, Burkina Faso, Kazakhstan, Turkey and Slovenia.