TDC VELOCE", 23G Two Dimensional Cutter - 20k CPM; Article Numbers: 9268.VIT23, 9311.23G01, 9311.23G02, 9311.23K01, 9311.23K02, 9311.23K03, 9623.G0201, 9623.G0202, 9623.K0201, 9311.23G01-00;

FDA Recall Z-2322-2026 · Posted 2026-06-05

Recall StatusOpen, Classified
Product CodeHQE
Date Initiated2026-04-30
Product Quantity314 boxes, totaling 1884 units

Reason for Recall

An increase in incidence in complaints of a loose outer knife and reports of the outer knife becoming loose during use.

Firm Action

On April 30, 2026 Urgent Medical Device Recall letters were sent to customers. Actions to be taken: Review this Product information as soon as possible with relevant members of your staff. 1. Pass this FSN as soon as possible on to all those who need to be aware of it, including other locations, organizations and/or customers who might have received the impacted Product(s) through you. 2. If you have the affected Products and the affected lots (see attachment below) in your inventory, remove and return them to DORC following the instructions in the attachment. 3. Follow the instructions on how to complete and submit the Field Action Acknowledgement Form that is provided with this FSN.

Root Cause

Process design

Distribution Pattern

Worldwide - US Nationwide and the countries of AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CZECHIA, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, HONG KONG, IRELAND, ITALY, JAPAN, NETHERLANDS, NEW ZEALAND, POLAND, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, TAIWAN, TURKEY.