Terminated

Evita 2 dura Critical Care Ventilator. Continuous (Respirator)

FDA Recall Z-2325-2008 · Posted 2008-09-20

Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeCBK
Date Initiated2008-07-02
Date Terminated2009-09-01
Product Quantity83 units

Reason for Recall

Audible power failure alarm was not enunciated at the required volume

Firm Action

The recalling firm issued an URGENT MEDICAL DEVICE RECALL letter dated 7/2/08 informing the consignees of the problem and the need to have the circuit board replaced.

Root Cause

Nonconforming Material/Component

Distribution Pattern

The products were shipped to medical facilities in AZ, CA, CO, FL, GA, IL, KY, MA, MI, MS, NJ, NY, PA, TN, TX, UT and WA.