IntelliVue Patient Monitor Product Usage: The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. The monitors are intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitors are intended for use by trained healthcare professionals in a hospital environment.
FDA Recall Z-2328-2016 · Posted None
Advertisement
Recalling FirmPhilips Electronics North America Corporation
Recall StatusTerminated
Product CodeMHX
Date Initiated2016-07-26
Date Terminated2020-07-14
Product Quantity50,495
Reason for Recall
If an affected Patient Monitor has been powered on continuously for several months, any displayed waveforms will contain outdated data and therefore fail to reflect the patients current condition.
Firm Action
Philips sent an Urgent Medical Device Correction letter dated July 20, 2016 to all affected customers. The Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Customers were informed that Philips will provide a software correction free of charge. For any further information or support concerning this issue, contact your local Philips representative at 1-800-722-9377.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution - US Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K161531IntelliVue Patient Monitors MP2, MP5, MP5T, MP5SC, and Multi-Measurement Module X2; IntelliVue Patient Monitors MP20, MP30, MP40, MP50, MP60; IntelliVue Patient Monitors MP70, MP80, MP90;IntelliVue Patient Monitors MX400, MX430, MX450, MX500, MX550, MX600;IntelliVue Patient Monitors MX700, MX800Philips Medizin Systeme Boeblingen GmbH · 2016-07-01K131872INTELLIVUE PATIENT MONITOR MX500, INTELLIVUE PATIENT MONITOR MX550Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · 2013-10-17
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.