Terminated

Puritan Bennett 980 Ventilator System, Universal (with neonatal functionality enabled) and Neonatal Models. Designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb).

FDA Recall Z-2329-2015 · Posted 2015-08-14

Recall StatusTerminated
Product CodeCBK
Date Initiated2015-07-16
Date Terminated2016-06-02
Product Quantity657 units

Reason for Recall

Reports in which tidal volumes reaching patients were lower than set tidal volumes in neonatal Volume Control Plus (VC+) Mode with active humidification. This situation may potentially lead to respiratory compromise if not recognized.

Firm Action

Consignees will be notified via letter on July 17, 2015. No products will be returned, the reconfiguration of affected ventilators will occur at customer sites.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution-US (nationwide) and the countries of Canada, United Arab Emirates, Argentina, Australia, Bahrain, Brazil, Chile, Colombia, Costa Rica, Egypt, Ireland, Israel, Jordan, Japan, Kuwait, Lebanon, Morocco, Mexico, Panama, Peru, Qatar, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Turkey, and Uruguay.