Triathlon Prim CEM FXD BPLT 7, Catalog No. 5520B700

FDA Recall Z-2329-2021 · Posted None

Recall StatusTerminated
Product CodeJWH
Date Initiated2021-07-01
Date Terminated2023-05-17
Product Quantity14 US; 2 OUS

Reason for Recall

The internal profile of the Triathlon Prim (Baseplate) is potentially oversized; there may be a witness mark (taper) at the bottom of the internal profile of the baseplate, which prevents the insert from locking on the baseplate.

Firm Action

On July 1, 2021, the firm notified customers via Urgent Medical Device Recall. The letter listed the hazards and harms potentially associated with the recalled product and also listed mitigation factors. Patients treated with an affected product should continue to be followed per the normal protocol established by his or her surgeon(s). There are no recommended changes to the frequency of the standard follow-up care protocol. It should be noted that the failure of the insert to fully seat, may not be evident on x-ray unless the insert has dislodged. Actions Needed: Our records indicate that you may have received the affected product(s). It is Strykers responsibility as the manufacturer to ensure that customers who may have received these affected products also receive this important follow up communication. We therefore request that you read this notice carefully and complete the following actions. 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list. Return the devices back to Stryker. Customers should complete and sign the included customer response form and return it to the recalling firm. Hospitals should contact their local sales office or Stryker sales representative directly for product returns and inventory questions. Customers who have further distributed the product should forward should notify downstream accounts and return all affected devices to the recalling firm. For questions pertaining to the recall, please contact Customer Service at (201) 831-5000, or email [email protected]

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states AR, CO, MN, MS, NJ, OK, PA, TX and the countries of China and Italy.